Agilent Technologies E1418A User Manual Page 103

  • Download
  • Add to my manuals
  • Print
  • Page
    / 190
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 102
TEST
The TEST subsystem provides commands used to run individual self-tests and to
query the results of previous self-tests. To run the complete set of self-tests, use the
*TST? common command.
Subsystem Syntax TEST
:NUMBer? <
number>
:TST
[:RESults]?
*TST?
TEST:NUMBer?
TEST:NUMBer? <number> performs a single internal self-test on the module and
returns a 16-bit integer indicating the results of the self-test.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
number>
numeric 1 to 75 none
Comments Following the completion of this command, all outputs are disabled and set to
either 0 A or 0 V.
If the module is in the wait-for-trigger state when this command is received, an
error is generated.
Complete descriptions of each test and specific codes returned in the Test Info
field are contained in the Agilent E1418A Service Manual.
A successful self-test returns 0.
A failure of a self-test returns a non-zero 16-bit integer. The integer returned has
a binary coding of the form:
BIT1514131211109876543210
Meaning Test Number A24 Test Info
The description of *TST provides a brief listing of the test numbers.
TEST
Chapter 3 Agilent E1418A SCPI Command Reference 101
Page view 102
1 2 ... 98 99 100 101 102 103 104 105 106 107 108 ... 189 190

Comments to this Manuals