Agilent Technologies E1418A User Manual Page 133

  • Download
  • Add to my manuals
  • Print
  • Page
    / 190
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 132
Software Trigger
Register
The Software Trigger Register provides a means to trigger the module when
software trigger has been selected as the trigger source in the Trigger
Control Register.
Address
Base + 0C
16
Bit 15141312111009080706050403020100
WRITE Software Trigger
Software Trigger: When software trigger is selected as the trigger source
in the Trigger Control Register, any write to this register generates the
trigger (the value written is not important).
Trigger Control
Register
The Trigger Control Register selects the sources, destinations, and modes
used to synchronize channels programmed for triggered operation.
Address
Base + 0E
16
Bit 15 141312111009080706050403020100
READ Trig
Syst En
Unused Trigger Out Trigger
Mode
Trigger Source
WRITE
Trig Syst En: Writing a one (1) to this field enables the trigger system.
Writing a zero (0) disables the trigger system.
Trigger Out: This field selects which VXIbus trigger line will be driven when
the trigger mode is set to Output. The trigger lines selected are shown below.
Line Out Selected
Bits
08 07 06
TTLTRIG0* 0 0 0
TTLTRIG1* 0 0 1
TTLTRIG2* 0 1 0
TTLTRIG3* 0 1 1
TTLTRIG4* 1 0 0
TTLTRIG5* 1 0 1
TTLTRIG6* 1 1 0
TTLTRIG7* 1 1 1
Trigger Mode: This field selects the operating mode for trigger-enabled
channels.
Trigger Mode
Bits
05 04
SYNC 0 0
OUTPUT 1 1
Appendix B Agilent E1418A Register-Based Programming 131
Page view 132
1 2 ... 128 129 130 131 132 133 134 135 136 137 138 ... 189 190

Comments to this Manuals

No comments