Appendix A
Agilent E1418A Specifications
DC Voltage Range:
±16 V
Resolution:
16 bits (488 µV steps)
Monotonic to 2.0 mV
Accuracy:
±(% of value + volts)
After calibration and at a temperature within ±5 °C of calibration temperature.
90 day: ±(0.05% + 3.0 mV)
Output Current:
Compliance Current : >20 mA, 0 to ±12 Volts,
derate linearly to 5 mA at ±16 Volts
Short Circuit Current: <60 mA
Differential Ripple and Noise:
< 2 mV rms, 20 Hz to 250 kHz into 10 k load
DC Current Range:
-0.02 Amps to +0.02 Amps
Resolution:
16 bits (610 nA steps)
Monotonic to 2.5 µA
Accuracy:
±(% of value + amps)
After calibration and at a temperature within ±5 °C of calibration
temperature.
90 day: ±(0.09% + 5.0 µA)
Output Voltage:
Compliance Voltage: ±12 V
Maximum Open Circuit Voltage: < 18 V
Differential Ripple and Noise:
< 2 µA rms, 20 Hz to 250 kHz into a 250 load