Adding 8 Channels Use this procedure to install an 8-Channel expansion kit (Agilent E1524A or
Agilent E1525A).
CAUTION Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic
discharge voltages as low as 50 volts. Disassemble and reconfigure only in a
static free work area. Minimize handling of assemblies and components. Keep
all assemblies and replacement parts in the original static free packaging.
These procedures should be performed by qualified service personnel at
approved static safe workstations.
Tools Needed T10 Torx Driver (one is provided in the kit).
Static Safe Workstation.
Static Wrist Strap.
Procedure Refer to Figures E-1 and E-2 during these procedures.
1. Remove the top cover. Refer to Figure E-1.
2. Remove the screw holding the sheet metal spacer and save the screw.
Remove and discard the sheet-metal spacer.
3. Place expansion board over the main board, component sides facing
each other. Line up the three electrical connectors and carefully press
into place. Install the hold down screw through the expansion board
and secure with a T10 Torx Driver. Refer to Figure E-2.
4. Replace the top cover.
5. Perform the adjustment procedures on the module. See Appendix D,
beginning on page 155.
Appendix E Configuration and Disassembly 169