CALibration:STORe
CALibration:STORe loads the stored non-volatile memory calibration constants
with the calibration constants currently in use by the module.
Parameters None.
Comments • This command sets the calibration constants in non-volatile memory if the CAL
Store Enable Jumper is properly set (see page 157). Non-volatile memory has
finite life, based upon the number of writes. Use this command sparingly.
• At power-on, calibration constants stored in non-volatile memory are loaded into
the module where they are used during output operations.
• When this command is received, the checksum is verified using the constants
currently in use by the module. If the checksum does not verify, the non-volatile
memory is not updated and an error is generated.
• Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 155,
for additional information about this command.
• Related Commands: CALibration:RESet
CALibration:TEST?
CALibration:TEST? performs an internal rough check of the calibration of each
channel and returns an integer indicating the channel calibration condition.
Calibration is checked to approximately ± 2%.
Parameters Returns: unsigned 16-bit integer (uint16).
Comments • Following the completion of this query all outputs are disabled and set to 0 Volts
or 0 Amps based on the channel mode.
• This query returns a 16-bit integer indicating the calibration status of each channel.
– Bits set to 0 indicate no calibration errors were found.
– Bits set to 1 indicate a channel with questionable calibration.
– Bits 0–15 correspond to channels 1–16, respectively.
• If the trigger system is waiting for a trigger when this command is received, an
error is generated.