CALibration:VALue:RESistance
CALibration:VALue:RESistance <value> enters a new measured value to be used
as the internal calibration resistor constant by the module.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
value> numeric (float 32) Must be > 0 ohms none
Comments The new calibration constant is used by the module but not stored in non-volatile
memory.
This command is used together with the CALibration:CONFigure:RESistance?
query to calibrate the internal current calibration resistor.
The internal calibration resistor should be calibrated before calibrating channels in
current mode. When calibrating at the CAL connections and using the internal
calibration resistor, the value need only be entered once to calibrate all channels.
If using an external resistor to calibrate current mode, you must calibrate at each
channel output. External resistances in the range of 50 to 500 are recommended.
Refer to Appendix D, Voltage/Current Output Adjustment”, starting on page 155,
for additional information about this command.
Related Commands: CALibration:STORe, CALibration:CONFigure:RESistance?
CALibration
70 Agilent E1418A SCPI Command Reference Chapter 3