  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 169
1st Value Returned Indicates an 8-channel or 16-channel configuration.
Bit 1514131211109876543210
Use Not Used Channels
Channels can have one of two values; all 1’s (decimal 7) for an 8-channel
configuration, and all 0’s (decimal 0) for a 16-channel configuration.
2nd Value Returned Indicates the type of Terminal Module (if any) installed. See page 77 for
additional information.
3rd Value Returned Indicates isolated or non-isolated plug-on modules for each channel.
Bit 1514131211109876543210
Use Ch16 Ch15 Ch14 Ch13 Ch12 Ch11 Ch10 Ch9 Ch8 Ch7 Ch6 Ch5 Ch4 Ch3 Ch2 Ch1
When a bit is set to a 0 an isolated plug-on is present. A bit set to a 1
indicates a non-isolated plug-on. For 8-channel devices, channels 9 through
16 bits are set to 1.
4th Value Returned Indicates the state of the V/I jumper on the terminal module as either current
or voltage when the P/J jumper is in the jumper selectable position. See
page 77 for additional information.
5th Value Returned Indicates the state of the channel output relay. See page 77 for additional
information.
6th Value Returned Indicates the state of the P/J jumper on the terminal module. See page 77
for additional information.
Changing
Configuration
Configuration changes are made using Agilent add-on kits. These
procedures support the following add-on kits:
Kit Part Number Description Use
Agilent E1523A 1-Channel Isolation
Plug-On Module
Change one or more channels
from Non-Isolated to Isolated
Output.
Agilent E1524A 8-Channel
Non-Isolated
Expansion Kit
Add 8 Channels of Non-Isolated
Output.
Agilent E1525A 8-Channel Isolated
Expansion Kit
Add 8 Channels of Isolated
Output.
168 Configuration and Disassembly Appendix E