TRIGger:SOURce
TRIGger:SOURce sets the trigger source for the module.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
source> discrete TTLT0, TTLT1, TTLT2, TTLT3, TTLT4,
TTLT5, TTLT6, TTLT7, BUS, HOLD,
IMM, or EXT
none
Comments The <source> parameter defines the following trigger conditions:
<source>Trigger
TTLT0–TTLT7 Sets the trigger source to one of the VXIbus backplane trigger lines.
BUS or HOLD Sets software trigger. A *TRG, TRIG:IMM, or GET will trigger the output.
IMMediate Places the trigger system in the wait-for-trigger state and then triggers the output.
EXTernal Sets the trigger source to the EXTernal trigger line on the terminal module.
If this command is received when the module is in the wait-for-trigger state, an
error is generated.
This command set the source for the trigger in to the module. The module can also
output a trigger on the VXIbus TTLTrg lines using the
OUTPut:TTLTrgn command.
Triggering is described in more detail on page 54.
Related Commands: TRIGger:SOURce?
*RST Condition: Trigger source set to HOLD.
TRIGger:SOURce?
TRIGger:SOURce? queries the trigger source for the module.
Parameters Returns: string.
Comments The query returns one of the following unquoted strings:
String Trigger Source
TTLTn VXIbus backplane trigger line n
BUS Software trigger
HOLD Trigger hold (reset state)
IMM Immediate trigger
EXT External trigger
TRIGger
Chapter 3 Agilent E1418A SCPI Command Reference 105
Page view 106