Agilent Technologies B1500A User's Guide Page 112

  • Download
  • Add to my manuals
  • Print
  • Page
    / 204
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 111
3- 34 Keysight B1500A User’s Guide, Edition 14
Installation
Mounting Connectors
To Connect GNDU Output
The GNDU output connection examples are shown in Table 3-7
Table 3-7 GNDU Output Connections
CAUTION Use Keysight 16493L GNDU cable to connect the GNDU to your connector plate or
test fixture. Do not use the triaxial cable. The GNDU is rated for up to 4.2 A, while
the maximum current rating of the triaxial cable is 1 A.
Kelvin connections (remote sensing) non-Kelvin connections
Use a low-noise coaxial cable (Keysight part
number 8121-1189) from the connector to the
prober, socket, or DUT as shown.
To cancel the effects of cable resistance, connect the
Sense line and Force line as close as possible to the
terminal of the DUT.
Short the Sense and Force on the connector as
shown below. Measurement data will include the
residual resistance of the connection wire.
Use AWG 22 single-strand insulated wire
(Keysight part number 8150-2639) from the
connector plate to the prober, socket, or DUT.
For a quick connection where measurement
accuracy is not critical, connect only Force to the
DUT, without shorting the Sense and Force. With
this connection, the measurement data will
include residual resistance from the connection
cable between the GNDU and the connector
plate.
to
DUT
Insulator
Triaxial connector
Plate
Coaxial cable
Force
Sense
Common
to
DUT
Wire
Insulator
Triaxial connector
Plate
Force
Sense
Common
Page view 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 203 204

Comments to this Manuals

No comments