3- 14 Keysight B1500A User’s Guide, Edition 14
Installation
Installing Accessories
Table 3-2 To Connect 16442B (B1500A-A5F)
Cables
16442B
Terminals
B1500A Terminals
Keysight 16493J Interlock Cable, 3 m or 1.5 m Intlk Interlock
Keysight 16493L GNDU Cable, 3 m or 1.5 m GNDU GNDU
Keysight 16493K Kelvin Triaxial Cable, 3 m
or 1.5 m, for the Kelvin connection
Up to 3 SMUs can be connected to the 16442B
by the Kelvin connection.
SMU1 SMU Force
SMU2 Sense
SMU3 SMU Force
SMU4 Sense
SMU5 SMU Force
SMU6 Sense
Keysight 16494A Triaxial Cable, 3 m or 1.5 m,
for the non-Kelvin connection
SMU1 SMU Force
SMU2 SMU Force
SMU3 SMU Force
SMU4 SMU Force
SMU5 SMU Force
SMU6 SMU Force
Keysight 16493P SPGU Cable, 3 m or 1.5 m PGU1 SPGU Output 1
PGU2 Output 2
Keysight N1300A CMU Cable
a
, 3 m or 1.5 m
a. To extend the four-terminal pair (4TP) connection to the 16442B inputs,
connect the Hcur and Hpot cables together and the Lcur and Lpot cables
together by using the BNC T-type adapter (female-female-male, Keysight
part number 1250-2405), then connect them to the 16442B inputs.
To extend the 4TP connection to the socket module inputs, connect the
cables to the PGU1, PGU2, VSU1, VSU2, VMU1, or VMU2 connector
individually, and connect the Hcur and Hpot terminals together and the
Lcur and Lpot terminals together at the socket module inputs.
It is ok to leave open the ground wire extended from the CMU cable.
PGU1 CMU Hcur/Hpot
PGU2 Lcur/Lpot