Keysight B1500A User’s Guide, Ed ition 14 4- 47
Specifications
EasyEXPERT Software
Automatic analysis function
On a graphics plot, the markers and lines can be automatically located using the auto
analysis setup. Parameters can be automatically determined using automatic
analysis, user function, and read out functions.
Analysis functions
Up to 20 user-defined analysis functions can be defined using arithmetic
expressions.
Measured data, pre-defined variables, and read out functions can be used in the
computation. The results can be displayed on the LCD.
Read out functions
Built-in functions for reading various values related to the marker, cursor, or line
Arithmetic functions
User functions
Up to 20 user-defined functions can be defined using arithmetic expressions.
Measured data and pre-defined variables can be used in the computation. The
results can be displayed on the LCD.
Arithmetic operators
+, -, *, /, ^, abs (absolute value), at (arc tangent), avg (averaging), cond
(conditional evaluation), delta, diff (differential), exp (exponent), integ
(integration), lgt (logarithm, base 10), log (logarithm, base e), mavg (moving
average), max, min, sqrt, trigonometric function, inverse trigonometric function,
and so on
Physical constants assigned to the keys
q: Electron charge, 1.602177E-19 C
k: Boltzmann constant, 1.380658E-23 J/K
(e): Dielectric constant of vacuum, 8.854188E-12 F/m
Numeric constants assigned to the keys
a (10
-18
), f (10
-15
), p (10
-12
), n (10
-9
), u or (10
-6
), m (10
-3
), k (10
3
), M (10
6
), G
(10
9
), T (10
12
), P (10
15
)
Page view 186