3- 12 Keysight B1500A User’s Guide, Edition 14
Installation
Installing Accessories
Installing Accessories
This section describes how to install the accessories.
“To Connect 16442B Test Fixture (B1500A-A5F)”
“To Connect Connector Plate”
“To Connect ASU”
“To Connect SCUU/GSWU”
“To Connect GNDU Adapter”
“To Connect MCSMU Connection Box”
WARNING To avoid electrical shock and instrument damage, turn the instrument off
before connecting or disconnecting measurement cable.
WARNING There are potentially hazardous voltages (± 200 V for HPSMU, and ± 100 V for
MPSMU/HRSMU) present at the Force, Sense, and Guard terminals. To
prevent electrical shock, do not expose these lines.
WARNING This instrument supports the Interlock function to prevent you from receiving
an electrical shock from high voltage (more than ± 42 V). If the Interlock
terminal is open, this instrument cannot apply high voltage more than ± 42 V.
If you do not use the 16442A/B test fixture, install an interlock circuit (see page
31) and connect it to the Interlock terminal.
CAUTION The GNDU can sink current of up to 4.2 A. Use Keysight 16493L GNDU cable to
connect the GNDU to a test fixture or a connector plate.
Do not use normal triaxial cable (Keysight 16494A) because the maximum current
rating of the cable is 1 A.
CAUTION Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the SMU.