Keysight B1500A User’s Guide, Ed ition 14 3- 17
Installation
Installing Accessories
To Connect ASU
Keysight E5288A Atto Sense and Switch Unit (ASU) adds the following functions
to the B1500A. To use the ASU, it must be connected to the SMU which is
connected together when the calibration is performed.
Atto ampere (aA) level current measurement capability
Automatic switching of ASU input resource connected to ASU output
NOTE For the installation of the ASU and the connection to the DUT interface (probe card,
manipulators, and so on), contact your favorite prober vender. The prober vender
will have the solutions. Dimensions of the ASU are 132 mm (W) 88.5 mm (H)
50 mm (D) excluding the connectors.
Figure 3-1 Image of ASU Installation
Required parts Keysight 16495K Connector plate
Keysight 16493M ASU cable
Keysight N1300A CMU cable or Keysight 16048D/E test leads
Keysight N1254A-108 Magnet stand
BNC cable, coaxial cable, manipulator, wire, and so on
To install 16495K Fix the 16495K plate to the shielding box or something that will cover the DUT
interface. See Keysight 16495 Installation Guide. The 16495K is the plate that has
the mechanism to block the light from the cable hole used to pass the cables in the
shielding box.
Shielding box
To B1500A/CMU
16495K 16495K
ASU
DUT interface
ASU
Prober
CMU Return
CMU Return