Keysight B1500A User’s Guide, Ed ition 14 2- 19
Introduction
Measurement Modules
HPSMU - High Power SMU
This section describes typical specification of the high power source/monitor unit
(HPSMU). Maximum four modules can be installed in one mainframe.
Maximum voltage, current, output power: ± 200 V, ± 1 A, 20 W
Minimum range: 2 V, 1 nA
Output/measurement value and resolution: see Table 2-1 to Table 2-4.
Figure 2-3 HPSMU Output and Measurement Ranges
200 200
100100
40
40
1000
50
125
1000
Voltage (V)
Current (mA)
20
500
500
20