Lab Exercise: Overview and Diagnostics
Diagnosis View
54
Note: Exceeding an EMF limit will not shut down your system or abort any
method or sequence. It simply changes the appearance of the EMF indicator.
When the EMF indicator is yellow, the user should go to the Diagnosis view,
click on the EMF tool and examine the EMF Memo Pad to find out what
preventive maintenance is due by comparing the Limit and the Actual values.
4) Return to the Diagnosis view.
Diagnosis Instrument Panel
In the left center of your Diagnosis view, a schematic of each instrument module
is shown. Information such as firmware revisions, the number of cycles a valve
has made, and the number of hours the lamp has burned can be accessed from
these graphical tools. A logbook, module tests, and module temperature history
are also available. In this section, you will explore these options.
1) Click on the pump diagram, and then select Update Variables Display.
Note that pump module information such as the Serial Number and
Firmware Revision is displayed below the diagram.
2) Click on the pump diagram again. This time, select Show Module Details.
A detailed schematic of the pump appears.
3) In a box to the right, you will see a symbol for the Logbook, an open book
(representing the method) and a thermometer. Click on the Logbook
symbol, then Maintenance Logbook Entry. When you perform routine
maintenance, enter the maintenance activity into the Maintenance Logbook.
If an instrument counter is associated with a maintenance activity, you will
be prompted to reset the counter. Cancel the Maintenance Logbook.
4) Again, click on the Logbook symbol, then Update Variables Display.
Below you will find access to the Error Logbook, Runs Logbook,