Agilent Technologies All Troubleshooting Guide Page 106

  • Download
  • Add to my manuals
  • Print
  • Page
    / 211
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 105
Lab Exercise: Quantification
Testing the Calibration Table
98
Testing the Calibration Table
1) Load the medium standard's data file, med.d. Go to the Report menu, then
Specify Report... or select the Specify Report Calculation and Print Style
tool.
2) Make the following selections:
Destination = Printer and Screen
Calculate = ESTD
Based on = Area
Report Style = Detail
Select chromatogram output.
3) OK the report specifications and save this to your method.
4) From the Report menu, select Print Report or select the Identify Peaks,
calculate Results & print Report tool.
5) Study the resulting report and modify retention time windows or qualifiers if
necessary. Remember to resave the method if you are satisfied with
report style.
Page view 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 210 211

Comments to this Manuals

No comments