Lab Exercise: Sequence Review and Reprocessing
Reprocess with Sequence Folder Method
111
Reprocess with Sequence Folder Method
You will reprocess a sequence after updating the Folder Method. This sequence
has two problems. First, calibrated compounds fell outside the method’s retention
time windows, and therefore, no amounts were calculated. The retention time
windows are set in the Calibration Settings.
In addition, the 4
th
level of the calibration is out of specifications and needs to be
dropped from the analysis. See the report below.