36 Data Organization
4 Workflow with Unique Folder Creation switched off
Working with Unique Folder Creation on or off?
Working with Unique Folder Creation on or off?
The new data concept as outlined in the previous chapters provides a number
of advantages:
Sequence data are not overwritten. Each sequence acquisition stores the
resulting data files in its own sequence container with unique name.
With the sequence container concept, the data are stored with all necessary
information needed for data analysis, i.e. copies of the sequence file and of
all methods employed with the sequence. These methods can be changed
with sequence specific input and do not influence the original master
method. The container concept thus strengthens the meaning of a sequence
as a set of data files and methods belonging together for result creation.
Data review and reprocessing are both available in Data Analysis view via
the Navigation Table.
The data container concept provides the optimal preconditions for the
ChemStation Integration with the Agilent Enterprise Content Manager
(ECM).
However, there may be situations where users may want to store their data as
in ChemStation B.01.03 or earlier and work according to the corresponding
workflows:
During method development it may be more convenient to have only one
method for both acquisition and data analysis to automatically have
changes available for future acquisition and reanalysis of already acqired
data.
Data from several acquisitions have to be in one folder, e.g. in case of partial
acquisition.
Customized macro solutions on a ChemStation system that have been
designed for older revisions may require the data, methods, or sequence to
be stored according to the old data organization scheme.
When ChemStation B.03.01 runs in a lab where there are also system still
running on ChemStation revisions B.01.03 or earlier, it may be more
convenient to use the same data organization mode on all systems.