Lab Exercise: Qualitative Data Analysis
Isoabsorbance Plot
41
selected in Data Analysis, that of the peak at 4.85 minutes. Convince yourself
that this is the case by moving the vertical cursor to a time of 4.85 minutes.
6) Now point the cursor at the horizontal line across your Isoabsorbance plot. As
you move this cursor up and down, watch the chromatogram in the lower
section of the screen. The black chromatogram changes as you change the
sample signal wavelength.
7) Change the cursor mode to Signal. To extract a signal corresponding to the
sample and reference wavelengths you determined earlier, you can simply
type the values you want in the box at the upper left. To access the reference
signal click in the check box next to Ref.
8) You can also select the sample and reference signals graphically. Press and
hold the CNTRL key while you move the sample signal band or the reference
signal band. You can keep the same bandwidth, but just adjust the position of
the band, by releasing the CNTRL key and then moving the signal band.
Note: If you want to start over, just exit the Isoabsorbance plot and redraw it
by selecting the Spectra menu, then Isoabsorbance plot. Don’t forget to
change the cursor mode to Signal.
9) When you are satisfied with your extracted signal, select the Copy button at
the lower left corner of the Isoabsorbance plot. Minimize the isoplot window
and view the extracted signal. Overlay the extracted signal with signal B.
You will have to select All Loaded Signals with the drop box.
Did you improve the signal to noise ratio of the peak?
10) You may continue to try other sample signals, bandwidths and references.
When you have finished, record the optimized values here.
Sample Signal__________
Sample Signal Bandwidth__________
Reference Signal__________
Reference Bandwidth__________
11) Close the IsoPlot.
If you have time, optimize the Diode Array signals for the 3rd and last major
peak in the acquisition data file. Record the results here and use in the next
laboratories.
Sample Signal__________
Sample Signal Bandwidth__________
Reference Signal__________
Reference Bandwidth__________