Lab Exercise: Quantification
Preparations
85
Preparations
Obtain the low, medium and high standards. Insert them into vial positions 1, 2
and 3 of the autosampler. Turn on the detector lamp. Prime your solvent delivery
system.
Priming the Agilent 1100 or 1200 Solvent Delivery System
1) If you are equipped with a vacuum degasser, make certain the degasser is on.
2) Ensure that the outlet tube is connected from the purge valve to the waste
container. Open the purge valve located on the pump module by turning the
knob counter clockwise several turns.
3) To prime the solvent delivery system, you should pump 100% at 5 mL/min
from each channel for several minutes. To accomplish this task, select the
following menu items: Instrument, Set up Pump...; or click on the pump set
up GUI in the system diagram (shown below), then Set up Pump....
4) Type in a flow of 5.000 mL/min and %B of 100. OK the dialog box.
Click