Agilent Technologies B1500A User's Guide Page 146

  • Download
  • Add to my manuals
  • Print
  • Page
    / 204
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 145
4- 6 Keysight B1500A User’s Guide, Edition 14
Specifications
Hardware
Ultra-fast IV measurement capability for the pulsed IV and transient IV such as
NBTI/PBTI, RTN, etc.
Waveform generation with 10 ns programmable resolution
Simultaneous high-speed IV measurement capability (200 MSa/s, 5 ns sampling
rate)
10 V peak-to-peak output
No load line effect accurate pulsed IV measurement by dynamic SMU
technology
Maximum Module Configuration
The B1500A can contain any combination of the following module. Then the total
power consumption of all SMU modules cannot exceed 84 W.
Up to 10 MPSMU
Up to 10 HRSMU
Up to 4 HPSMU
Up to 4 MCSMU
•One MFCMU
Up to 5 HV-SPGU
Up to 5 WGFMU
When one or more WGFMU modules are installed in the B1500A mainframe, the
following table applies. Multiply the values given below by the number of installed
modules of that type and add the products together. The sum of the products must be
less than or equal to 59 for the configuration to be permissible.
Maximum Voltage between Common and Ground
± 42 V
MPSMU 2
HRSMU 2
HPSMU 14
MCSMU 5
MFCMU 7
HV-SPGU 12
WGFMU 10
Page view 145
1 2 ... 141 142 143 144 145 146 147 148 149 150 151 ... 203 204