Agilent Technologies 16048A Specifications Page 19

  • Download
  • Add to my manuals
  • Print
  • Page
    / 104
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 18
16034H Test Fixture
Terminal Connector: 4-Terminal Pair, BNC
DUT Connection: 2-Terminal
Dimensions (approx.):
120(W) x 50(H) x 70(D) [mm]
Weight (approx.): 200 g
Additional Error:
f: [MHz]
15
Up to 110 MHz (4-Terminal Pair) SMD
Type of Error Impedance
Proportional Error 0.5 x (f/10)
2
[%]
Open Repeatability 5 + 500 x (f/10) [nS]
Short Repeatability 10 + 13 x (f/10) [mΩ]
Description: This test fixture is designed for impedance
evaluations of array-type SMD. The minimum SMD size
that this fixture is adapted to evaluate is 1.6(L) x 0.8(W)
[mm]. Since the tip of the measurement electrodes are
very thin and the device holder is extremely flat, the
device can be shifted and the measurement electrodes can
contact the each elements of the array-type component.
Applicable Instruments: 4263B, 4268A, 4279A*, 4284A*,
4285A, 4288A, 4294A, E4980A, E4981A
* denotes the instrument is obsolete.
Frequency: DC to 110 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below
Furnished Accessories:
Compensation and Measurement: Open and short
compensations are recommended before measurement.
When measuring above 3 MHz, load compensation is also
recommended. Open compensation is performed by
separating the high and the low electrodes from each
other. The separation should be equivalent in size to the
DUT’s width. Short compensation is performed by
placing the high and low electrodes in contact together.
Load compensation is performed by using the furnished
100 Ω SMD chip resistor. After performing open, short
and load compensations, the DUT is inserted into the test
fixture. Refer to the 16034G figures to see how
compensation and measurement are performed.
Description P/N Qty.
Case for 100 Ω SMD Resistance 1540-0692 1
100Ω Chip Resistor 0699-2488 10
Operation Manual 16034-90012 1