Agilent Technologies 16048A Specifications Page 37

  • Download
  • Add to my manuals
  • Print
  • Page
    / 104
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 36
42842B Bias Current Test Fixture
DUT Connection: 2-Terminal
Bias Voltage Output connector: BNC(f)
Dimensions (approx.):
237(W) x 173(H) x 235(D) [mm]
Weight (approx.): 3000 g
Basic Measurement Accuracy:
2% (< 1 kHz), 1% (1 kHz)
For detailed information, refer to the operation
manual of 42841A.
33
Up to 110 MHz (4-Terminal Pair) DC Bias Accessories
Description: This test fixture is designed for use with
the 42841A. It is connected directly to the bias current
output terminals of the 42841A. The following features
ensure accurate and safe DC bias measurements:
Clear plastic cover is provided for safe measurements.
Opening the cover decreases the voltage generated by
back-e.m.f. (electromotive force) to a level below 40 V
within 0.1 second.
Heat-sensitive switch prevents DUT from overheating.
Also equipped with DC bias voltage monitoring.
Applicable Instrument: E4980A with Option E4980A-002,
4284A* with Option 4284A-002
* denotes the instrument is obsolete.
Frequency: 20 Hz to 2 MHz
Maximum DC Bias Current: 40 A
Operating Temperature: 5°C to 45°C
DUT Size: Smaller than 80(W) x 80(H) x 80(D) [mm]
Furnished Accessories:
Compensation and Measurement: Short compensations
is recommended before measurement. Short compensation
is performed by shorting the measurement terminals
together with the furnished shorting plate. After compen-
sation, connect the DUT to the measurement terminals
and close the test fixture cover.
Description P/N Qty.
Protection Caps N/A 2
Shorting Plate 42842-00607 1
Operation and Service Manual 42842-90002 1
Page view 36