Agilent Technologies 16048A Specifications Page 66

  • Download
  • Add to my manuals
  • Print
  • Page
    / 104
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 65
62
16094-65000 Probe Test Fixture
Terminal Connector: 7 mm
DUT Connection: 2-Terminal
Electrical Length: 23.2 mm
Weight (approx.): 25 g
Additional Error: See figure below
Up to 3 GHz (7 mm) Probe
Description: This probe test fixture is used along with the
7 mm cable (P/N 8120-4779). Together they provide the
capability to perform in-circuit measurements (printed cir-
cuit patterns, the input/output impedance of circuits, etc.).
Applicable Instrument: 4294A + 42942A, (4287A, 4395A
with Opt. 4395A-010 + 43961A, 4396B with Opt. 4396B-010
+ 43961A, E4991A)**
When used with 16085B: 4263B, 4268A, 4279A*, 4284A*,
4285A, 4288A, E4980A, E4981A
* denotes the instrument is obsolete.
** applicable in a limited frequency range.
Frequency: DC to 125 MHz
Maximum DC Bias Voltage: ±40 V
Operating Temperature: 0°C to 55°C
DUT size: See figure below.
Compensation and Measurement: The compensation
with following 3 steps is recommended.
1. Perform open/short/load calibration at the end of 7 mm
coaxial extension cable before connecting the
16094-65000 to the cable.
2. After connecting the 16094-65000 to the tip of the
coaxial cable, enter the electrical length value of the
16094-65000, 23.2 mm, to the instrument.
3. Perform open and short compensations at the tip of the
16094-65000. The open compensation is performed by
separating the high and the low electrodes from each
other. Short compensation is performed by shorting the
probe. To short the probe, it is recommended to use a
shorting device with gold-plated surfacing (which
provides stable contact resistance). The figures in this
page shows the images of open and short conditions.
Necessary Accessories:
The operating note of this product is not furnished as stan-
dard. It is downloadable from Agilent website (www.agi-
lent.com). Search the web using the keyword “16094A”.
Open compensation
Short compensation
L: 1 mm ~ 15 mm