Agilent Technologies 16048A Specifications Page 34

  • Download
  • Add to my manuals
  • Print
  • Page
    / 104
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 33
30
Up to 110 MHz (4-Terminal Pair) DC Bias Accessories
16065A 200Vdc External Voltage
Bias Fixture
Terminal Connector: 4-Terminal Pair, BNC
DUT Connection: 4-Terminal
External Bias Input connector: High Voltage BNC(f)
Dimensions (approx.): 180(W) x 120(H) x 200(D) [mm]
Cable Length (approx.): 40 cm
Weight (approx.): 1500 g
High Voltage BNC(f) connector for external bias input
BNC(f) connector for voltage monitor output
LCR meter with 16065A
16065C 40Vdc External
Voltage Bias Adapter
Terminal Connector: 4-Terminal Pair, BNC
External Bias Input connector: BNC(f)
Dimensions (approx.): 160(W) x 50(H) x 150(D) [mm]
Cable Length (approx.): 210 mm
Weight (approx.): 450 g
LCR meter with 16065C
Description: This test fixture makes it possible to mea-
sure a DUT with up to ±200 V DC bias. The same mod-
ules of 16047A/D can be used to allow measurements of
axial/radial lead components.
Applicable Instruments: 4263B, 4268A, 4279A*,
4288A, E4981A (4284A*, 4285A, 4294A, E4980A)**
* denotes the instrument is obsolete.
** applicable in a limited frequency range.
Frequency: 50 Hz to 2 MHz
Maximum DC Bias: ±200 V
Blocking Capacitor of 5.6 µF is connected in with the Hc
terminal.
Operating Temperature: 0°C to 55°C
DUT Size: See the 16047A figure with module sizes.
Furnished Accessories:
Compensation and Measurement: Open, short and
load compensations are recommended before measure-
ment. Short compensation is performed by shorting the
contacts of the test fixture with a shorting plate as
described for the 16047A. Load compensation is performed
by inserting a known standard device. After performing
open, short and load compensations, the DUT is connected
to the test fixture.
Description P/N Qty.
Module For Axial Lead 16061-70022 1
Module For Radial Lead mounting
16061-70021 1
on fixture
Module For Short Radial Lead 16047-65001 1
Shorting Bar 5000-4226 1
Operation and Service Manual 16065-90011 1
Description: This adapter is designed to operate specifi-
cally with the 4263B, 4268A, 4288A and the E4981A. By
connecting an external DC voltage source to this adapter,
a bias voltage of up to ±40 V can be supplied to a DUT.
The DUT can be inserted by connecting any direct attach-
ment 4-Terminal Pair test fixture to the adapter.
Applicable Instruments: 4263B, 4268A, 4288A, E4981A
Frequency: 100 Hz to 1 MHz
Maximum DC Bias: ±40 V
Blocking Capacitor of 100 µF is connected in series with
the Hc terminal.
Operating Temperature: 0 °C to 55 °C
Applicable Fixtures: 16034E/G/H, 16044A,
16047A/D/E,
16048A/B*/D/E, 16089A/B/C/D/E
* denotes the instrument is obsolete.
Furnished Accessories:
Compensation and Measurement: Open and short com-
pensations are recommended before measurement. Short
compensation is performed by shorting the contacts of the
test fixture that is in use. After performing open and short
compensations, the DUT is connected to the test fixture.
Description P/N Qty.
Operation and Service Manual 16065-90020 1
Page view 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 103 104

Comments to this Manuals

No comments