Agilent Technologies 16048A Specifications Page 26

  • Download
  • Add to my manuals
  • Print
  • Page
    / 104
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 25
Up to 110 MHz (4-Terminal Pair) Other Components
16089A Large Kelvin Clip Leads
Terminal Connector:
4-Terminal Pair, BNC
DUT Connection: 4-Terminal
Cable Length (approx.):
0.94m (from connector to clip's tip)
Weight (approx.): 300 g
Additional Error: The additional error is negligible
when compared to the instrument's accuracy.
Description: This test fixture makes it possible to mea-
sure odd-shaped components that cannot be measured
with conventional fixtures. It is equipped with two insu-
lated Kelvin clips.
Applicable Instruments: 4263B, 4268A, (4284A*, 4285A,
4288A, 4294A, E4980A, E4981A)**
* denotes the instrument is obsolete.
** applicable in a limited frequency range.
Frequency: 5 Hz to 100 kHz
Maximum Voltage: ±42 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size:
See figure below
Furnished Accessories:
Compensation and Measurement: Open and short
compensations are recommended before measurement.
For open compensation, do not connect the Kelvin clips to
anything. Short compensation is performed by holding a
shorting plate with the Kelvin clips. After performing
open and short compensations, the DUT is held with the
Kelvin clips.
16089B Medium
Kelvin Clip Leads
Terminal Connector: 4-Terminal Pair, BNC
DUT Connection: 4-Terminal
Cable Length (approx.):
0.94m (from connector to clip's tip)
Weight (approx.): 300 g
Additional Error: The additional error is
negligible when compared to the instrument's
accuracy.
Description P/N Qty.
Operation and Service Manual 16089-90020 1
Description: This test fixture makes it possible to mea-
sure odd-shaped components that cannot be measured
with conventional fixtures. It is equipped with two insu-
lated Kelvin clips.
Applicable Instruments: 4263B, 4268A, (4284A*,
4285A, 4288A, 4294A, E4980A, E4981A)**
* denotes the instrument is obsolete.
** applicable in a limited frequency range.
Frequency: 5 Hz to 100 kHz
Maximum Voltage: ±42 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below
Furnished Accessories:
Compensation and Measurement: Open and short
compensations are recommended before measurement.
For open compensation, do not connect the Kelvin clips to
anything. Short compensation is performed by connecting
the Kelvin clips together. After performing open and short
compensations, the DUT is held with the Kelvin clips.
Description P/N Qty.
Operation and Service Manual 16089-90020 1
22
Page view 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 103 104

Comments to this Manuals

No comments