Agilent Technologies B1500A User's Guide Page 153

  • Download
  • Add to my manuals
  • Print
  • Page
    / 204
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 152
Keysight B1500A User’s Guide, Ed ition 14 4- 13
Specifications
Hardware
Table 4-5 Voltage range, resolution, and accuracy (high resolution ADC)
Table 4-6 Current range, resolution, and accuracy (high resolution ADC)
Range
Force
resolution
Measurement
resolution
Force accuracy
±(%+mV)
a
a. ±(% of output/measured value + offset value in mV)
Measurement accuracy
±(%+mV)
a
Maximum
current
±2 V 100 V2 V ±(0.018 + 0.4) ±(0.01 + 0.14) 1 A
±20 V 1 mV 20 V ±(0.018 + 3) ±(0.009 + 0.9) 1 A
±40 V 2 mV 40 V ±(0.018 + 6) ±(0.01 + 1) 500 mA
±100 V 5 mV 100 V ±(0.018 + 15) ±(0.012 + 2.5) 125 mA
±200 V 10 mV 200 V ±(0.018 + 30) ±(0.014 + 2.8) 50 mA
Range
Force
resolution
Measurement
resolution
Force accuracy ±(%+A+A)
a
a. ±(% of output/measured value + fixed offset in A + proportional offset in A), Vo is the output
voltage in V.
Measurement accuracy
±(%+A+A)
a
Maximum
voltage
±1 nA 50 fA 10 fA ±(0.1+3E-13+Vo1E-15) ±(0.1+2E-13+Vo1E-15) 200 V
±10 nA 500 fA 10 fA ±(0.1+3E-12+Vo1E-14) ±(0.1+1E-12+Vo1E-14) 200 V
±100 nA 5 pA 100 fA ±(0.05+3E-11+VoE-13) ±(0.05+2E-11+Vo1E-13) 200 V
±1 A 50 pA 1 pA ±(0.05+3E-10+VoE-12) ±(0.05+1E-10+Vo1E-12) 200 V
±10 A 500 pA 10 pA ±(0.05+3E-9+Vo1E-11) ±(0.04+2E-9+Vo1E-11) 200 V
±100 A 5 nA 100 pA ±(0.035+15E-9+VoE-10) ±(0.03+3E-9+Vo1E-10) 200 V
±1 mA 50 nA 1 nA ±(0.04+15E-8+Vo1E-9) ±(0.03+6E-8+Vo1E-9) 200 V
±10 mA 500 nA 10 nA ±(0.04+15E-7+VoE-8) ±(0.03+2E-7+Vo1E-8) 200 V
±100 mA 5 A 100 nA ±(0.045+15E-6+VoE-7) ±(0.04+6E-6+Vo1E-7)
b
b. 200 V (Io 50 mA), 100 V (50 mA < Io 125 mA), 40 V (125 mA < Io 500 mA),
20 V (500 mA < Io 1 A), Io is the output current.
±1 A 50 A1 A ±(0.4+3E-4+VoE-6) ±(0.4+15E-5+Vo1E-6)
Page view 152
1 2 ... 148 149 150 151 152 153 154 155 156 157 158 ... 203 204

Comments to this Manuals

No comments