4- 26 Keysight B1500A User’s Guide, Edition 14
Specifications
Hardware
Table 4-14 Example of calculated C/G measurement accuracy
Frequency
Measured
Capacitance
C Accuracy
a
a. The calculation examples are specified under the conditions Dx = 0.1, 1 PLC integration time,
30 mVrms test signal level, and at four-terminal pair port of MFCMU.
Measured
Conductance
G Accuracy
a
5 MHz 1 pF ± 0.61 % 3 S ± 192 nS
10 pF ± 0.32 % 31 S ± 990 nS
100 pF ± 0.29 % 314 S ± 9 S
1 nF ± 0.32 % 3 mS ± 99 S
1 MHz 1 pF ± 0.26 % 628 nS ± 16 nS
10 pF ± 0.11 % 6 S ± 71 nS
100 pF ± 0.10 % 63 S ± 624 nS
1 nF ± 0.10 % 628 S ± 7 S
100 kHz 10 pF ± 0.18 % 628 nS ± 11 nS
100 pF ± 0.11 % 6 S ± 66 nS
1 nF ± 0.10 % 63 S ± 619 nS
10 nF ±
0.10 % 628 S ± 7 S
10 kHz 100 pF ± 0.18 % 628 nS ± 11 nS
1 nF ± 0.11 % 6 S ± 66 nS
10 nF ± 0.10 % 63 S ± 619 nS
100 nF ± 0.10 % 628 S ± 7 S
1 kHz 100 pF ± 0.92 % 63 nS ± 6 nS
1 nF ± 0.18 % 628 nS ± 11 nS
10 nF ± 0.11 % 6 S ± 66 nS
100 nF ± 0.10 % 63 S ± 619 nS