4- 32 Keysight B1500A User’s Guide, Edition 14
Specifications
Hardware
Table 4-16 Pulse/DC output voltage and accuracy
ALWG (arbitrary linear waveform generator) function
Arbitrary linear waveform generator (ALWG) mode
Output complex waveform per one channel of HV-SPGU
Define multi-level pulse and multi-pulse waveform including open state
pulse with ALWG GUI editor
Sequential pulse waveform from user-defined pulse waveform
1024 points per one channel
Item Value Condition
Output voltage (Vout) -20 V to +20 V 50 load
-40 V to +40 V open load
Accuracy
a
a. At 1 s after completing transition.
± (0.5 % + 50 mV) open load
Amplitude resolution 0.2 mV (± 10 V range) 50 load
0.8 mV (± 40 V range)
0.4 mV (± 10 V range) open load
1.6 mV (± 40 V range)
Output connectors SMA
Source impedance
50
b
b. Typical 1 %)
Short circuit current
800 mA peak (400 mA average
c
)
c. Specification condition: N 0.2 + Itotal < 3.0 A where, N is the number
of installed HV-SPGUs, and Itotal is DC current output by all modules
(including HV-SPGUs)
Overshoot/pre-shoot/ringing
d
d. Following the specified condition with transition time.
± (5 % + 20 mV) 50 load
Output limit Monitoring over current limit