Agilent Technologies B1500A User's Guide Page 176

  • Download
  • Add to my manuals
  • Print
  • Page
    / 204
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 175
4- 36 Keysight B1500A User’s Guide, Edition 14
Specifications
Hardware
The V Monitor terminal outputs a buffered signal equal to 1/10 of the Output
terminal voltage (into a 50 load).
WGFMU to RSU cable length
The WGFMU and RSU are connected by a special composite cable. The
following configurations are available.
•3 m
•5 m
1.5 m
2.4 m + connector adapter + 0.6 m
4.4 m + connector adapter + 0.6 m
Note: The connector adapter is used when routing the cable through the probers
connector panel.
ALWG function: See Table 4-21.
Trigger output
Level: TTL
Trigger width: 10 ns
Generated synchronously with ALWG waveform.
•Software
Instrument library for WGFMU control
NBTI and general-purpose EasyEXPERT Application Tests
Sample programs (NBTI and general-purpose measurement using WGFMU
and RTS data analysis)
WGFMU supported prober vendors
Cascade Microtech
Suss MicroTec
Vector Semicon
Note: The maximum number of installable RSUs for a given prober depends
upon the available space. Please contact your local sales representative for
details on connecting and mounting the WGFMU and RSU.