Keysight B1500A User’s Guide, Ed ition 14 4- 41
Specifications
Hardware
Table 4-23 Minimum timing parameters for voltage measurement
1
Voltage applied to DUT 5 V 10 V
Applied voltage condition
Recommended minimum pulse width
a
a. Recommended minimum pulse width = settling time + recommended minimum measurement
window.
105 ns 130 ns
Voltage measurement condition Measurement Range 5 V 10 V
Recommended minimum measurement window 20 ns 20 ns
Settling time
b
b. The time until the measured value settles to within ± 0.6 % of the final result value after the out-
put voltage is changed from the initial value (0 V). Minimum rise/fall time of 70 ns for 10 V, or
30 ns for 5 V is recommended for minimizing overshoot.
85 ns 110 ns
Noise (rms)
c
c. RMS noise measured over the recommended minimum measurement window.
1.4 mV 1.4 mV
1. Measurement conditions: The DUT is a resistive load between 1 k and 10 M. The capaci-
tance of the cable between the RSU and the DUT is 20 pF. Voltage is applied to the DUT by a
channel of WGFMU/RSU, and voltage measurement is performed by the same channel. (PG
mode for 5 V, Fast IV mode for 10 V)