Agilent B1500 Users Guide, Edition 7 2- 75
Introduction
Specifications
General Specifications
Temperature range
Operating +5 °C to +40 °C
Storage -20 °C t
o +60 °C
•Humi
dity
Operating 20 % to 70 %RH, non-condensing
Storage 10 % to 90 %RH, non-condensing
Altitude
Operating 0 to 2,000 m (6,561 ft.)
Storage 0 to 4,600 m (15,092 ft.)
Power requirement: 90 to 264 V, 47 to 63 Hz
Maximum volt-amps (VA): 900 VA
Regulatory compliance
EMC IEC61326-1:+A1 / EN61326-1:+A1
AS/NZS 2064.1
Safety CSA C22.2 No.1010.1-1992
IEC61010-1:+A2 / EN61010-1:+A2
UL3111-1:1994
Certification: CE, CSA, NRTL/C, C-Tick
•Dimensions
B1500A: 420 mm (W) × 3
30 mm (H) × 575 mm (D)
N
1301A-100 SCUU: 148 mm (W) × 75 mm (H) × 70 mm (D
)
N1301A-200 GSWU: 33.2 mm (W) × 41.5 mm (H) × 3
2.8 mm (D)
E5288A ASU: 132 mm (W) × 88.5 mm (H) × 50 mm (D)
B
1531A RSU: 45.2 mm (W) × 70 mm (H) × 82 mm (D)
16440A
Selector: 250 mm (W) × 50 mm (H) × 27
5 mm (D)
16445A Selector adapter: 250 mm (W) × 50 mm (H) × 26
0 mm (D)
•Weight
B1500A (empty) 20 kg, B1510A 2.0 kg, B1511A 1.0 kg, B1517A 1.2 kg,
E1520A 1.5 kg, E15
25A 1.3 kg, E1530A 1.3 kg, E1531A 0.13 kg, E5288A 0.5
kg, N1301A-100 0.8 kg, N1301A-200 0.1 kg, 16440A 1.1 kg, 16445A 1.0 kg