Agilent B1500 Users Guide, Edition 7 9-5
If You Have a Problem
When You Operate B1500A
To Simplify the Connections
If you want to simplify the measurement connections, use the triaxial cables instead
of the kelvin triaxial cables, and connect the triaxial cables between the Force
terminals and the test fixture or the connector plate. Then the Sense terminals must
be opened.
NOTE The B1500A has the Force and Sense terminals to make the kelvin connection. And
the kelvin triaxial cables are effective to extend the measurement terminals near by
the device under test. Connecting the Force and Sense lines together at the device
terminals can minimize the measurement error caused by the residual resistance of
the connection cables. The kelvin connection is effective for the low resistance
measurement and the high current measurement. Also it is one of the conditions to
satisfy the measurement specifications of the B1500A.