9-64 Agilent B1500 User’s Guide, Edition 7
If You Have a Problem
Error Codes
3312 Specified ALWG measurement interval time is out of absolute limits.
Check the measurement interval time and set the correct value.
The
value must be 10 ns to 1.34217728 s, in 10 ns resolution.
3313 Specified ALWG measurement instruction code is invalid.
Check the measurement event setting and set the correct values.
3314 Specified ALWG range change instruction code is invalid.
Check the range event setting and
set the correct values
.
3315 Specified ALWG measurement count is out of ab
solute limits.
Check the measurement averaging tim
e and set the correct value. The
value must be 0, or 10 ns to 0.020971512 s, in 10 ns resolution.
3316 Specified ALWG measurement count is greater than measurement
in
terval.
Check the measurement averaging tim
e and set the correct value. The
value must less than or equal to the measurement interval time.
3317 Specified slot is invalid.
Check the slot number and set the correct value. The slot number must
be 1 to 10.
3318
Specified module channel is invalid.
Check the channel number and
set the correct value.
3319
Output delay is out of absolute limits.
Check the output delay and set the correct value. The value must be
-50 ns to
50 ns, in 625 ps resolution.
3320 Measurement delay is out of absolute limits.
Check the measurement delay and set the correct val
ue. The value must
be -50 ns to 50 ns, in 625 ps resolution.
3321 VM/IM measurement mode is invalid.
Check the measurement mode and set the corr
ect value.
3322 Voltage measurement range is invalid.
Check the voltage measurement range and set the correct value.
3323 Current measurement range is invalid.