Installation
Requirements
Requirements
This section contains information on:
“Power Requirements”
“Operating Environment”
“Storaging/Shipping Environment”
“Ventilation Requirements
“Power Cable”
Power Requirements
The B1500A can operate from any single-phase AC power source supplying 90 to
264 V at 47 to 63 Hz. The maximum power consumption is 900 VA.
Operating Environment
The B1500A is specified to operate within the following environmental conditions:
Temperature: 5 °C to 40 °C
Humidi
ty: 20 % to 70 % RH, non-condensing
Altitude: 0 to 2,000 m (6,561 ft.)
Storaging/Shipping Environment
The B1500A is specified to store/ship within the following environmental
conditions:
Temperature: -20 °C to 60 °C
Humidi
ty: 20 % to 90 % RH, at 40 °C, non-condens
ing
Altitude: 0 to 4,600 m (15,092 ft.)