Agilent Technologies B1500A User's Guide Page 159

  • Download
  • Add to my manuals
  • Print
  • Page
    / 628
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 158
Agilent B1500 Users Guide, Edition 7 3- 29
Installation
Mounting Connectors
To Make an Interlock Circuit
The interlock circuit is designed to prevent electrical shock when a us
er touches the
measurement terminals.
CAUTION You must install an interlock circuit on a shielding box to prevent hazardous
voltages when the door of the shielding box is open.
Figure 3-8 shows the pin assignments of the interlock connector that should be
mounted on a connector plate or test fixture.
Figure 3-8 Interlock Connector Pin Assignments
WARNING Potentially hazardous voltages may be present at the Force, Guard, and Sense
terminals when the interlock terminals are shorted.

,QWHUORFN
VZLWFK

/('
:LULQJVLGHYLHZ
3OXJVLGHYLHZ
Page view 158
1 2 ... 154 155 156 157 158 159 160 161 162 163 164 ... 627 628

Comments to this Manuals

No comments