2- 32 Agilent B1500 User’s Guide, Edition 7
Introduction
Measurement Units
B1530A WGFMU
This section describes typical specification of the waveform generator/fast
measurement unit (WGFMU) for Agilent B1500A. Maximum five modules can be
installed in one mainframe.
Number of channels: 2 channels per module
Function:
Voltage output and current or voltage sampling measurement with minimum
sampling interv
al of 5 ns
Current measurement range: 1 μA, 10 μA, 10
0 μA, 1 mA, 10 mA
Operation mode: PG mode, Fast IV mode, DC mode
•PG mode:
ALWG (arbitrary linear waveform generator) voltage output and voltage
m
easurement (
VFVM). The output voltage will be divided by the internal 50 Ω
resistor and the load impedance.
Output level: -5 V to 5 V
Mi
nimum pulse width: 100 ns
•Fast IV mode:
ALWG voltage output and current or vo
ltage measurement (VFIM o
r VFVM).
Output level: 0 to -10 V, 0 to 10
V, -5 V to 5 V
Minimum pulse width: 300 ns
•DC mode:
DC voltage output and current or vol
tage measurement (VFIM o
r VFVM).
Output level: 0 to -10 V, 0 to 10
V, -5 V to 5 V
Figure 2-8 WGFMU Simplified Circuit Diagram
507
V
PG mode
A
V
ALWG
Ch1
RSU output
Fast IV mode
507
V
PG mode
A
V
ALWG
Ch2
RSU output