3- 36 Agilent B1500 User’s Guide, Edition 7
Installation
Mounting Connectors
Figure 3-14 ASU Output non-Kelvin Connection
The ASU inputs can be connected to the instruments
by using a control cable, a
triaxial cable, and two coaxial cables. The cables can be connected to the ASU
inside the shielding box through the 16495K plate.
The ASU outputs should be connected to the DUT by using coaxial cables and wires
as shown above.
Wh
en a SMU is used for the measurements, the Kelvin paths/force lines will work
as same as Ta
ble 3-8. Then the sense line is not extended to the DUT.
When a four-terminal pair (4TP) instrument is used, the se
nse lines can be ignored.
And a force line is used for the high signal, and the other one is used for the low
signal. For the accurate impedance measurements, a wire must be connected
between the ASU#1 CMU Return and the ASU#2 CMU Return.
If the 4TP instrument is not used, you can ignore the CMH, CML, CMU Return, and
the DUT1 high
and low terminals shown above.