5- 24 Agilent B1500 User’s Guide, Edition 7
Classic Test Definition
I/V-t Sampling
Base Hold Time Hold time of the Base value output until the Source value
output, in seconds. 0 to 655.35 s, resolution 0.01 s.
Stop Condition This area defines the sampling measurement stop condition. See “Stop Condition”
on page 7-18 for this function.
Enable/Disable Enables or disables the stop condition.
Enable Delay Delay time. in second. This is the time from starting sampling
measurement to
enabling this function.
Name Name of measurement data or user function to monitor for stop
condition.
Val of Event.
Threshold Threshold value at which to stop sampling measurement.
Th of Event.
Event Event for stop condition.
Val > Th True if Name parameter value is greater
than Thresh
old value.
Val < Th True if Name parameter value is less than
Threshold value.
|V
al| > |Th| True if absolute Name parameter value is
greater than absolute T
hreshold value.
|Val| < |Th| True if absolute Name parameter value is
less than abso
lute Threshold value.
No. of Events Target value of the count the event occurs (
true). When the
count of true is this value, sampling is immediately stopped.
Constants This area defines the constant voltage/current source.
Unit Source monitor unit (SMU) used for the constant source
V Name Variable name for the constant voltage output or measurement
data
I Name Va
riable name for the constant current output or
measurement
data
Mode Output mode, V (voltage), or I (current)
Base, Source Base or Source value, in V or A.
0 to ±100 V (HRSMU/MPSMU) or ±200 V (HPSMU)
0 to ±0.1 A (HRSMU/MPSMU) or ±1 A (HPSMU)