Agilent B1500 Users Guide, Edition 7 6- 33
Application Test Definition
GPIB I/O
GPIB I/O
This component provides the setup editor that contains the following entry fields.
This component is used to control the GPIB interface. For the fields to enter a
variable name, only the variable name defined in the test definition can be entered.
CAUTION Do not set the GPIB address of the B1500A to the Address parameter of this
component even if this is on the Desktop EasyEXPERT software. Operation of the
B1500A is not guaranteed.
Write String Sends the specified string data to the specified GPIB device.
Address GPIB address of the target device
String String value to send
Terminator Terminator, DEFAULT, EOI, CR/LF+EOI, LF+EOI, CR/LF,
LF
, or NONE (not
specified)
Write Value Translates the specified numeric data to the string data, and sends it to the specified
G
PIB device.
Address
GPIB address of the target device
Format Format of Write Value. Follow Agilent T&M Programmers
Toolkit.
Val
ue0 Numeric value or string value to send
Value1 Numeric value or string value to send
Value2 Numeric value or string value to send
Value3 Numeric value or string value to send
Terminator Terminator, DEFAULT, EOI, CR/LF+EOI, LF+EOI, CR/LF,
LF
, or NONE (not
specified).