9-4 Agilent B1500 User’s Guide, Edition 7
If You Have a Problem
When You Operate B1500A
ASU is not Detected
If Agilent B1500A/EasyEXPERT cannot detect ASU (atto sense/switch unit), turn
B1500A off, check the cable connection, and reboot B1500A/EasyEXPERT.
For the installation and connection of ASU, see “To Connect ASU” on page 3-17.
NOTE Connect ASU to dedicated HRSMU
The specifications are satisfied and guaranteed for the exclusive combination of the
ASU and the HRSMU. So confirm the serial number of the ASU and connect it to
the dedicated HRSMU properly.
External GPIB Devices are not Detected
If Agilent B1500A/EasyEXPERT cannot detect external GPIB devices, do the
following troubleshooting.
1. Turn B1500A and external GPIB devices off, check the GPIB cable connection,
and reboot the dev
ices and B1500A/EasyEXPERT.
2. If the problem still remains, confirm if B1500A is set to the system controller.
See “To Enable System Controller” on page 3-10.
3. If the problem still remains, confirm if B1500A is connected to USB0. Open
Agilent Co
nnection Expert, check Instrument I/O on this PC. The USB0 in this
area must show the information as shown below.
USB0
+ B1500A(USB0::xxxx::x::xxxx::x::INSTR)
+ agb1500a
+ UsbDevice1
If the problem still remains, perform Agilent B1500 system recovery. To perform
recovery, see “B1500A System Recovery” on page 9-19.