Agilent B1500 Users Guide, Edition 7 7- 35
Function Details
SPGU Module
In the example in Figure 7-19, SPGU1 and SPGU2 are used. The channels output
Pattern2 twice and Pattern1 three times in one sequence, although the two patterns
for SPGU1 are the sam
e.
The SPGU channels simultaneously start the sequence output. The sequence will be
repeated in the sa
me way as for the pulse period. The channels continue the
sequence output until the specified time elapses or the repeat count reaches the
specified number of sequences.
NOTE Setup delay time between the different patterns
When the pattern is changed to a pattern of a differing index, a 50 ns setup delay will
always be inserted.
There is no delay between patterns of the same index.