Agilent B1500 Users Guide, Edition 7 7- 65
Function Details
Auto Power Off Function
Auto Power Off Function
If an abnormal voltage or current is detected in Agilent B1500, Agilent B1500
automatically turns off to prevent damage.
If this function is activated, Agilent B1500 is turned off, but the Standby switch is
left ON.
Perform the following steps to turn Agilent B1500 on again.
1. If Windows is running, end all tasks, and shutdown Windows.
2. Open the measurement terminals.
3. Press the Standby switch to set it to the OFF position.
4. Remove the power cable from the B1500 rear panel.
5. Wait for 10 seconds or more.
6. Connect the power cable from the B1500 rear panel.
7. Press the Standby switch to set it to the ON position.
Abnormal voltage or current can be caused by incorrect connections of Agilent
B
1500 to the tes
t device, excess voltage or current input, momentary power loss,
defects of the instrument, and so on.
If no cable is connected to the measurement terminals of
Agilent B1500, and the line
voltage is correct, Agilent B1500 may have a defect. Contact the nearest Agilent
Technologies Service Center for assistance.