Chapter 9 133
Saving and Recalling (File Management)
Saving and Recalling File
9. Saving and Recalling
(File Management)
Saving state for each channel into a register
For the active channel, when you want to save the instrument state specific to that channel
only into one of registers A to D, use the following command:
:MMEM:STOR:CHAN on page 354
Recalling an instrument state saved in a register can reproduce it as the state of the active
channel. To recall a register, use the following command:
:MMEM:LOAD:CHAN on page 346
NOTE It is possible to recall a file from a different channel where it was saved.
The contents in the registers are lost when you turn OFF the power. You can delete (clear)
the contents of all the registers using the following command.
:MMEM:STOR:CHAN:CLE on page 354
Saving measurement data
Measurement data (in a formatted data array) can be saved in the file with CSV (Comma
Separated Value) format.
To save measurement data on a file, use the following command:
:MMEM:STOR:FDAT on page 356
Executing the above command will save the measurement data of the active trace.Note that
the data save using the above command cannot be recalled from the E5061A/E5062A.
Saving the images on the LCD screen
Images displayed on the LCD screen can be saved on a file in the bitmap (.bmp) or portable
network graphics (.png) format.
To save screen image on a file, use the following command:
:MMEM:STOR:IMAG on page 357
Executing the above command will save the screen image when the command is invoked.
NOTE Note that this gives different result than operation from the front panel where screen image
is saved at the time [Capture] key is pressed.