Agilent Technologies E5071C User's Guide Page 316

  • Download
  • Add to my manuals
  • Print
  • Page
    / 551
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 315
316 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:DISP:TABL
Syntax :DISPlay:TABLe[:STATe] {ON|OFF|1|0}
:DISPlay:TABLe[:STATe]?
Description Turns ON/OFF the display of the window selected with the :DISP:TABL:TYPE
command.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":DISP:TABL ON"
20 OUTPUT 717;":DISP:TABL?"
30 ENTER 717;A
Related commands :DISP:TABL:TYPE on page 317
Equivalent key
[Sweep Setup] - Edit Segment Table
[Marker] - Marker Table
[Analysis] - Limit Test - Edit Limit Line
[Macro Setup] - Echo Window
NOTE When performing the operation from the front panel, you select the window and turn
ON/OFF the display at the same time.
Description
ON or 1 Turns ON the display.
OFF or 0 (preset value) Turns OFF the display.
Page view 315
1 2 ... 311 312 313 314 315 316 317 318 319 320 321 ... 550 551

Comments to this Manuals

Vikiehs 04 Jan 2024 | 14:20:06

urenrjrjkvnm