Chapter 13 461
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:STAT:QUES:LIM:NTR
Syntax :STATus:QUEStionable:LIMit:NTRansition <numeric>
:STATus:QUEStionable:LIMit:NTRansition?
Description Sets the value of the negative transition filter of the Questionable Limit Status Register.
Parameters
Note that bit 5 to 15 cannot be set to 1.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":STAT:QUES:LIM:CHAN1:NTR 16"
20 OUTPUT 717;":STAT:QUES:LIM:CHAN1:NTR?"
30 ENTER 717;A
Related commands :STAT:QUES:LIM? on page 455
:STAT:QUES:LIM:PTR on page 462
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Value of the negative transition filter
Range 0 to 65535
Preset value 0
Resolution 1