Chapter 13 257
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:MARK:MATH:FST
Syntax :CALCulate{[1]|2|3|4}[:SELected]:MARKer:MATH:FSTatistics[:STATe] {ON|OFF|1|0}
:CALCulate{[1]|2|3|4}[:SELected]:MARKer:MATH:FSTatistics[:STATe]?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), turn on and off the filter statistics.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:MARK:MATH:FST ON"
20 OUTPUT 717;":CALC1:MARK:MATH:FST?"
30 ENTER 717;A
Related commands :CALC{1-4}:MARK:MATH:FST:DATA? on page 258
Equivalent key
[Marker Fctn] - RF Filter Stats
Description
ON or 1 Turns ON the filter statistics.
OFF or 0
(preset value)
Turns OFF the filter statistics.