Agilent Technologies E5071C User's Guide Page 99

  • Download
  • Add to my manuals
  • Print
  • Page
    / 551
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 98
Chapter 6 99
Analyzing Data
Notch search
6. Analyzing Data
Notch search
The E5061A/E5062A provides the notch search function. This function searches for points
that meet the conditions on the trace, initiated at the marker position and sweeping the trace
rightward, and calculates the following items.
R Bandwidth( )
R Center frequency ( )
R Q value ( )
R Loss (response value at the marker position)
Where, is the right-hand cutoff point frequency, and is the eft-hand cutoff
point frequency.
Setting the notch definition value
The notch search function finds a point whose response value is different, by the amount
defined as the notch definition value, than the response value at the marker position, and
identifies that point as the cutoff point.
To set the notch definition value, use the following command:
:CALC{1-4}:MARK{1-10}:NOTC:THR on page 277
Reading out the notch search result
Once you have moved the marker to the appropriate position using the marker search
function or some other function, you can retrieve the notch search result using the
following command:
:CALC{1-4}:MARK{1-10}:NOTC:DATA? on page 276
Also, you can use the following command to control whether to display the notch search
result on the LCD:
:CALC{1-4}:MARK:NOTC on page 261
NOTE You can retrieve the notch search result regardless of whether the marker display and the
notch search result display is on/off.
BW high low=
cent
high low+
2
---------------------------=
Q
cent
BW
-----------=
high low
Page view 98