374 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COEF?
Syntax :SENSe{[1]|2|3|4}:CORRection:COEFficient[:DATA]? {ES|ER|ED|EL|ET|EX},<numeric
1>, <numeric 2>
Description Returns the calibration coefficient for specified channel.
Parameters
If ES, ER, or ED is used, the response port and the stimulus port must be the same, while
EL, ET, or EX is used, the response port and the stimulus port must be different.
Query response {numeric 1},…,{numeric NOP×2}<newline><^END>
Where NOP is the number of measurement points and n is an integer between 1 and NOP.
Example of use
10 DIM A(1:201)
20 OUTPUT 717;":SENS1:CORR:COEF? EL,1,2"
30 ENTER 717;A(*)
Equivalent key No equivalent key is available on the front panel.
<numeric>
ES Source match
ER Reflection tracking
ED Directivity
EL Load match
ET Transmission tracking
EX Isolation
numeric 1 Response port
numeric 2 Stimulus port
Description
{numeric n×2-1} Real part of the data (complex number) at the n-th measurement
point.
{numeric n×2} Imaginary part of the data (complex number) at the n-th
measurement point.