Agilent Technologies E5071C User's Guide Page 222

  • Download
  • Add to my manuals
  • Print
  • Page
    / 551
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 221
222 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:FUNC:DOM
Syntax :CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:DOMain[:STATe] {ON|OFF|1|0}
:CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:DOMain[:STATe]?
Description For channel 1 (:CALC1) to channel 4 (:CALC4), sets whether to use an arbitrary range
when executing the analysis with the :CALC{1-4}:FUNC:EXEC command.
When the trace coupling is off, the active trace is the target to be set.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:FUNC:DOM ON"
20 OUTPUT 717;":CALC1:FUNC:DOM?"
30 ENTER 717;A
Related commands :CALC{1-4}:FUNC:EXEC on page 225
:CALC{1-4}:FUNC:DOM:COUP on page 223
:CALC{1-4}:FUNC:DOM:STAR on page 224
:CALC{1-4}:FUNC:DOM:STOP on page 225
Equivalent key No equivalent key is available on the front panel.
Description
ON or 1
Specifies an arbitrary range
*1
.
OFF or 0 (preset value) Specifies the entire sweep range.
*1.Use the :CALC{1-4}:FUNC:DOM:STAR command and the
:CALC{1-4}:FUNC:DOM:STOP command to specify a range.
Page view 221
1 2 ... 217 218 219 220 221 222 223 224 225 226 227 ... 550 551

Comments to this Manuals

Vikiehs 04 Jan 2024 | 14:20:06

urenrjrjkvnm